.AixLib.Fluid.FixedResistances.Validation.PlugFlowPipes.Data.BaseClasses.PipeDataULg

Information

This record contains information of an aproximately 15 minutes long test bench carried out at the University of Liège.

Data structure

Column 1: Time in s

Column 2: Mass flow rate in kg/s

Column 3: Outlet pipe temperature in °C

Column 4: Outlet water temperature in °C

Column 5: Inlet pipe temperature in °C

Column 6: Inlet water temperature in °C

Test procedure

Before to perform a test, the water city network is pushed inside the approximately 39 meter long studied pipe during about 10 minutes to be sure that it is at the same temperature. During this time period, valves V3 and V1 are opened, the boiler is off and the valve V2 is closed.

Then, the valve V1 is closed and the valve V2 and V3 are opened. The boiler is started to reach the setpoint hot water temperature. When the temperature setpoint is achieved, data starts to be recorded, the valve V1 is opened and the valve V2 is closed at the same time to supply the studied pipe in hot water. After the outlet pipe temperature is stabilized, the boiler is shut off.

During the test, the ambient temperature is equal to 18°C and the mass flow rate is set to 1.245 kg/s.

Test bench schematic

Schematic of test rig at ULg

Notice: length are approximate

Pipe characteristics


Generated at 2024-11-23T19:25:52Z by OpenModelicaOpenModelica 1.24.2 using GenerateDoc.mos